MEDDEV 2.12/1
 |
 |
MEDDEV 2.12/1February 1998 rev. 3
GUIDELINES
ON A MEDICAL DEVICES VIGILANCE SYSTEM
The present Guidelines are part of a set of Guidelines relating to questions of
application of EC-Directives on medical devices. They are legally not binding. The
Guidelines have been carefully drafted through a process of intensive consultation of the
various interested parties (competent authorities, Commission services, industries, other
interested parties) during which intermediate drafts were circulated and comments were
taken up in the document. Therefore, this document reflects positions taken by
representatives of interested parties in the medical devices sector.
Note : this document is a revision of an earlier document
published in March 1993 as MEDDEV. 3/93 rev. 2.
These Guidelines describe an adverse incident notification and evaluation system, to be
known as the Medical Devices Vigilance system.
CONTENTS
Page
1. INTRODUCTION 4
2. FOR WHOM THESE GUIDELINES ARE WRITTEN 5
3. PURPOSE OF THE VIGILANCE SYSTEM 5
4. GENERAL PRINCIPLES 6
5. TYPES OF INCIDENT TO BE REPORTED BY
MANUFACTURERS 7
6. MAKING AND RECEIVING AN INITIAL REPORT 10
7. PROCEDURE FOLLOWING THE INITIAL REPORT 13
8. OUTCOME OF AN INVESTIGATION AND FOLLOW-UP 15
9. REGISTER OF KEY TERMS 19
APPENDICES
1. LIST OF COMPETENT AUTHORITIES FOR VIGILANCE 22
2. DEFINITIONS 24
3. SUGGESTED INCIDENT REPORT FORMAT 25
4. SIMPLIFIED FLOW CHART - ILLUSTRATION OF INCIDENTS TO BE REPORTED 31
5. EXAMPLES OF INCIDENTS TO REPORT 32
6. SIMPLIFIED FLOW CHART ILLUSTRATING THE REPORTING OF RECALLS 34
7. RELEVANT EXTRACTS FROM THE DIRECTIVES 35
8. SUGGESTED COMPETENT AUTHORITY REPORT FORMAT 37
1. INTRODUCTION
1.1.1 These Guidelines describe a system for the notification and evaluation of adverse
incidents to be known as the Medical Devices Vigilance system. They are intended to
facilitate the uniform application and implementation of the Directive for Active
Implantable Medical Devices (AIMD) and the Directive for Medical Devices (MDD). In
particular, Article 8 (AIMD) and Article 10 (MDD) outline the obligations of Member States
upon receipt of incident reports, from manufacturers or other sources, concerning any
medical device which carries the CE marking.
These Guidelines also give guidance on those Annexes of the Directives which oblige the
manufacturer to report certain types of incident to Competent Authorities.
These Guidelines are not enforceable by law.
Relevant extracts from both Directives are provided in Appendix 7.
A diagrammatic summary of the system is given in Appendix 9.
1.1.2 These Guidelines cover the activities of :
- the Commission,
- Competent Authorities,
- Notified Bodies,
- manufacturers (including their authorised representatives and persons responsible for
placing on the market, see Article 14 of the MDD),
- users and others concerned with the continuing safety of medical devices.
For the purposes of Medical Devices Vigilance, Member States are represented by the
Competent Authorities listed in Appendix 1.
1.1.3 These Guidelines cover the action to be taken once the manufacturer or Competent
Authority receives information concerning an incident. Information on incidents which
should be reported under the Vigilance system may come to the attention of manufacturers
via the systematic procedure to review experience gained from devices in the
post-production phase, or by other means (see Annexes 2, 4, 5, 6, 7 of MDD). The term
"post-marketing surveillance" as referred to in Annex 2, 4, 5 in AIMD has the
same meaning as the aforementioned "systematic procedure..".
These Guidelines make no recommendations on the structure of the systems by which
manufacturers gather information concerning the use of devices in the post-production
phase.
1.1.4 This medical devices vigilance system takes precedence over information systems
to be installed under the requirements of the Council Directive 92/59/EEC of 29/6/1992 on
General Product Safety
1.2 USER AND OTHER INCIDENT REPORTING
SYSTEMS
1.2.1 Member States may wish to supplement reports received from manufacturers under
the Vigilance system with reporting from other sources. Member States should adopt
administrative measures to ensure that the pertinent manufacturers are informed without
delay of reports meeting the criteria set out later in these Guidelines (see para 6.4).
1.2.2 In order to enhance the efficacy of the Medical Device Vigilance system,
Competent Authorities should encourage the reporting of adverse events by the user. Such
reports may be made either directly to the Competent Authority, or to the manufacturer, or
to both depending on National practice.
1.3 APPLICABILITY
1.3.1 These Guidelines refer to incidents occurring within the Member States of the
European Community and all other States within the European Economic Area (EEA) with
regard to:
* devices which carry the CE-mark ;
and
* devices which do not carry the CE-mark, where such incidents lead to corrective
action relevant to CE-marked devices.
Corrective action includes, but may not be confined to: device recall; issue of
advisory notice; additional surveillance/modification of devices in use; modification to
future device design, components or manufacturing process; modification to labelling or
instructions for use.
These Guidelines do not apply to devices under clinical evaluation or investigation -
see Annex 7 of AIMD and Annex 10 of MDD.
1.3.2 If incidents which occur outside the EEA lead to corrective action relevant to
CE-marked devices which are offered for sale or are in use within the EEA, then
manufacturers should notify the relevant Competent Authorities.
1.3.3 These Guidelines are intended to be applicable equally to the Directive for
Active Implantable Medical Devices and the Medical Devices Directive. The procedures are
intended to be the same for both Directives, with respect to the Vigilance system.
2 FOR WHOM THESE GUIDELINES ARE WRITTEN
2.1 MANUFACTURERS
These Guidelines apply to manufacturers placing medical devices on the market in
accordance with the AIMD and the MDD. The definition of the term "manufacturer"
is given in Appendix 2.
2.1.2 Manufacturers should ensure that these Guidelines are made known to their
authorised representatives within the EEA, persons responsible for placing devices on the
market and any other agents authorised to act on their behalf for purposes related to
Medical Devices Vigilance, so that the manufacturers' responsibilities may be fulfilled.
2.2 COMPETENT AUTHORITIES
These Guidelines cover Competent Authorities' responsibilities, in particular under
Article 2 of AIMD and MDD, and article 8 of AIMD and Article 10 of MDD.
2.3 NOTIFIED BODIES
Notified Bodies may be consulted by Competent Authorities or manufacturers following
incidents involving, for example, medical devices for which they have provided
attestations leading to a CE mark.
2.4 HEALTH-CARE ORGANISATIONS AND
PERSONNEL
Member States should ensure that organisations and individuals involved in purchasing
medical devices and in the provision of health-care are aware that their co-operation is
vital in providing the first link in the Vigilance chain. This includes organisations and
individuals responsible for providing calibration and maintenance for medical devices.
3 PURPOSE OF THE VIGILANCE SYSTEM
3.1.1 The purpose of the Vigilance system is to improve the protection of health and
safety of patients, users and others by reducing the likelihood of the same type of
adverse incident being repeated in different places at different times. This is to be
achieved by the evaluation of reported incidents and, where appropriate, dissemination of
information which could be used to prevent such repetitions, or to alleviate the
consequences of such incidents.
3.1.2 The Vigilance system is intended to allow data to be correlated between Competent
Authorities and manufacturers and so facilitate corrective action earlier than would be
the case if data were collected and action taken on a State by State basis.
3.1.3 Whilst the manufacturer has the responsibility for taking any action necessary,
Competent Authorities should also monitor the effectiveness of the manufacturers'
follow-up on reported incidents.
The Competent Authority should take any further action that may be necessary to
supplement the actions of the manufacturer.
3.1.4 Once corrective (or other) action is identified, hospital administrators, medical
practitioners and other health-care professionals, and user representatives responsible
for the maintenance and the safety of medical devices, can take the necessary steps. Such
steps should, where practicable, be taken in co-operation with the manufacturer.
3.1.5 Competent Authorities may also monitor experience with devices of the same kind
(for instance, all defibrillators or all syringes), but made by different manufacturers.
They may then be able to take measures applicable to all devices of that kind. This could
include, for example, initiating user education or suggesting re-classification.
4. GENERAL PRINCIPLES
4.1.1 Information held by Competent Authorities in connection with the Vigilance System
is to be held in confidence, as defined by the relevant Articles of the Directives (AIMD
15 and MDD 20). In order to achieve the purpose of the Vigilance system, any incident
report should be available on request, and in confidence, to the other Competent
Authorities (see paragraph 7.2).
4.1.2 The act of reporting an incident to a Competent Authority is not to be construed
as an admission of liability for the incident and its consequences. Written reports may
carry a disclaimer to this effect. (see paragraph 6.2)
4.1.3 The initial report on an incident under the Vigilance system is made by the
manufacturer to the Competent Authority for recording and evaluation (see
paragraph 6). Each initial report should lead to a final report (see paragraph 8.6),
but not every initial report will lead to a corrective action.
4.1.4 The manufacturer should ensure that their authorised representative within the
EEA, persons responsible for placing devices on the market and any other agents authorised
to act on their behalf for purposes relating to Medical Devices Vigilance, are kept
informed of incident reports as appropriate
The manufacturer should consider informing official distributors etc as appropriate
during the procedure. This does not affect the right of the manufacturer to determine the
person authorised to be the principle contact point for purposes of relating to Medical
Devices Vigilance.
4.1.5 It is recommended that manufacturers inform their Notified Body of those
incidents which may affect the certification provided by that Notified Body. However, it
remains the role of the Competent Authority to monitor the investigation being carried out
by the manufacturer.
4.1.6 Depending on the outcome to the investigation, any information necessary for the
prevention of further incidents (or the limitation of their consequences) should be
disseminated (see paragraph 8.4 and 8.5).
5. TYPES OF INCIDENTS TO BE
REPORTED BY MANUFACTURERS TO COMPETENT AUTHORITIES
5.1 REQUIREMENTS OF THE ANNEXES
Extracts from the Annexes of the AIMD and of the MDD which define what should be
reported by the manufacturer to Competent Authorities are given in Appendix 7. Although
the wording of these Annexes to the two Directives is different, the interpretation given
in these Guidelines is nevertheless the same.
For example, the Directive for Medical Devices includes the word "serious" as
a qualification of "deterioration in his state of health". In these Guidelines,
the Directive for AIMDs is interpreted as though the word "serious" were
present.
5.2 DECISION PROCESS ON WHAT A MANUFACTURER SHOULD REPORT
The manufacturer should assess the following points when deciding whether an incident
should be reported to a Competent Authority:
- the type of incident (or potential incident) - paragraph 5.4;
- whether any medical device may have been involved which was made by that manufacturer
or under his authority;
- whether the incident was caused (wholly or partly), or could have been caused, by the
device or by shortcomings in the information supplied with the device - paragraph 5.5.
The same considerations apply to a Competent Authority's decision whether to inform a
manufacturer of an incident reported via a User Reporting or other system -see paragraph
6.4.
5.3 ACCESS TO THE DEVICE SUSPECTED TO BE
INVOLVED IN THE INCIDENT
A manufacturer may consult with the user on a particular incident before a report has
been made to the Competent Authority (see paragraph 6.1). The manufacturer may also wish
to have access to the device said to be involved in the incident for the purpose of
deciding whether the incident should be reported to the Competent Authority. Such access
may be affected by the requirements of National Law, and may also be at the discretion of
the user or health-care facility concerned.
If the manufacturer gains access to the device, and his initial assessment (or cleaning
or decontamination process) will involve altering the device in a way which may affect
subsequent analysis, then the manufacturer should inform the Competent Authority before
proceeding. The Competent Authority may then consider whether to intervene.
5.4 GUIDELINES ON TYPES OF INCIDENTS TO BE
REPORTED
The following paragraphs describe the types of
incidents which a manufacturer should report to the Competent Authority.This is
illustrated by the simplified flow chart and the examples of such incidents, given in
Appendices 4 and 5.
In assessing the type of incident, the manufacturer should consult with the medical
practitioner involved or other health-care professional wherever practicable.
Incidents which need to be reported are defined in the Directives as follows:
5.4.1 Those which led to a death;
5.4.2 Those which led to a serious deterioration in the state of health of a patient,
user or other person.
A serious deterioration in state of health can include:
- life-threatening illness or injury;
- permanent impairment of a body function or permanent damage to a body structure;
- a condition necessitating medical or surgical intervention to prevent permanent
impairment of a body function or permanent damage to a body structure.
Note: The interpretation of the term serious is not easy, and should be made
in consultation with a medical practitioner wherever possible. Many points may need
consideration, for example:
- whether a risk was foreseeable and clinically acceptable in view of potential patient
benefit;
- whether the outcome was adversely affected by a pre-existing condition of the
patient.
In cases of doubt on this issue, it is suggested that there should be a pre-disposition
to report rather than not to report.
5.4.3 Those which might have led to death or serious deterioration in health
Not all incidents which should be reported involve a death or serious deterioration in
health which actually occurred. The non-occurrence of such a result might have been due to
other fortunate circumstances or to the intervention of health-care personnel.
It is sufficient that:
- an incident associated with a device happened, and
- the incident was such that, if it occurred again, it might lead to death or serious
deterioration in health.
OR
- an examination of the device or the information supplied with the device indicated
some factor (eg a deterioration in characteristics or performance, or a shortcoming in the
information) which could lead to an incident involving death or serious deterioration in
health;For the purposes of these Guidelines, such potential incidents are to be known as
"near incidents".
For a near incident to be reported, a possible direct link with the device, or with
shortcomings in the information supplied, should be clearly established.
5.5 THE INCIDENT OR NEAR INCIDENT AND THE DEVICE OR THE INFORMATION SUPPLIED WITH THE
DEVICE
The following paragraphs describe the
characteristics of the device, or of the information supplied with the device, which may
be associated with an incident which should be reported.
In assessing the link between the device and the incident or near incident, the
manufacturer should take account of:
- the opinion, based on available evidence, of health-care professionals;
- the results of the manufacturer's own preliminary assessment of the incident;
- evidence of previous, similar incidents;
- other evidence held by the manufacturer.
5.5.1 Malfunction or deterioration in the characteristics and/or performance of a
device.
A malfunction is a failure of a device to perform in accordance with its intended
purpose (see Appendix 2) when used in accordance with the manufacturer's instructions.
This includes single fault conditions (e.g. component failures) which cause, or could
cause, an injury. (The EN 60601-1 definition of single fault is reproduced in Appendix 2
Definitions).
It does not include single fault conditions for which the manufacturer has made
provision (in accordance with relevant standards) and which do not cause an injury, or
potential injury.Deterioration of the device also includes problems caused by unpredicted
biological effects related to the device.
Deterioration does not include the normally expected ageing of implant materials or
depletion of batteries which were predicted in the information supplied with the device.
5.5.2 A device which shows no malfunction or deterioration, but nevertheless has a
characteristic which could lead to an incident should be reported as a near incident.
5.5.3 Inaccuracies in the instruction leaflet, or instructions for use include
omissions and deficiencies.
The terms "instruction leaflet" and "instructions for use" include
all information provided on or with the device, such as instruction materials, or user and
maintenance manuals.
An example of an omission is a failure to warn of a side effect which may be produced
by the device working within specification; an example of a deficiency is a lack of
clarity which leads, or could lead to, an injury.
Omissions do not include the absence of information which should generally be known by
the intended user.
Any inaccuracy in the instructions which caused, or could cause, misuse or incorrect
maintenance or adjustment should be reported.
5.5.4 Reference to the above considerations may be made in the report, or should be
kept on file by the manufacturer in the case a decision not to report.
5.6. TIMESCALE FOR THE INITIAL REPORTING
OF AN INCIDENT OR NEAR INCIDENT
The report should be made as soon as possible
commensurate with determining whether the incident falls within the guidance discussed
above. The times given below are the maximum elapsed times for determining the relevant
facts and making an initial report.
The time runs from the manufacturer first being informed of the incident, to the
relevant Competent Authority receiving the notification from the manufacturer.
Incidents 10 days
Near incidents 30 days
5.7 SYSTEMATIC RECALLS
5.7.1 The Directives require any technical or medical reason for the systematic recall
of a device to be notified by the manufacturer to a Competent Authority. The Term
"withdrawal" used in the AIMD is interpreted in the same way.
The term "recall" is defined in EN46001 (see Appendix).
Removals from the market for purely commercial reasons are not included.
A simplified flow chart illustrating the types of recalls to be reported is given at
Appendix 6.
5.7.2 The manufacturer should issue advisory notices when implementing recalls. Copies
of advisory notices should be sent to the Competent Authorities of the countries to which
they are applicable, and for devices in Class II or Class III, the Competent Authority in
the State where the Notified Body is situated and which made the attestation which led to
the CE marking being attached to the device. Manufacturers should consider sending copies
of Advisory Notices to Competent Authorities under cover of a report which takes the same
structure as the Final Report (see para 8.1.3). Notification to Competent Authorities
should be made before or at the same time as the notices are sent to the relevant users.
The terms "advisory notice" and "Recall" are defined in EN46001 (see
Appendix 2).
6 MAKING AND RECEIVING AN INITIAL
REPORT BETWEEN MANUFACTURER AND COMPETENT AUTHORITY
6.1 COMPETENT AUTHORITY TO WHICH AN
INITIAL REPORT SHOULD BE MADE
6.1.1 In general, the report should be to the
Competent Authority in the country of occurrence of the incident, with the following
provisos or exceptions :
i) In the case of an incident involving an implant which occurs in a Member State other
than the Member State where the implant was performed, the above principle still applies.
In addition, the manufacturer should copy the report to the Competent Authority of the
State where the implant was performed, if known.
ii) Reports on incidents concerning devices in Class II or Class III occurring in
countries outside the EEA and which result in corrective action, should be made to the
Competent Authority in the State where the Notified Body is situated and which made the
attestations which led to the CE marking being attached to the device.
iii) Reports on incidents concerning Class I devices which occur outside the EEA
and which result in corrective action, should be made to the Competent Authority of the
Member State in which the manufacturer, or the person responsible for placing on the
market, has made his notification within Article 14 of MDD.
The list of Competent Authorities is in Appendix 1. The list is correct at the time of
writing, but there may have been changes, additions or deletions since.
6.1.2 Where appropriate, manufacturers should notify their authorised representative,
persons responsible for placing on the market and any other agents authorised to act on
their behalf of incidents reported under the Vigilance System.
Manufacturers should inform their Notified Body of those incidents reported under the
Vigilance System which relate to the conformity assessment as carried out be the Notified
Body.
6.2 DETAILS TO BE INCLUDED IN INITIAL
REPORT
A recommended format for the initial report by the
manufacturer to the Competent Authority is given in Appendix 3.
6.2.1 The report should include the following details as appropriate:
- manufacturer's name (and the name of the authorised representative within the EEA,
where relevant), address, contact point, telephone number, Fax.
- the date when the incident came to the knowledge of the manufacturer;
- medical device kind, commercial name, catalogue number model, serial/batch/lot
number, software version;
- identification number of the Notified Body involved in the conformity assessment
procedure (if any), and the date(s) of the attestation(s);
- associated devices and/or accessories involved in the incident (if known);
- details of the incident (to the extent known) including date and patient or user
outcome;
- current location of device involved in the incident, if known;
- contact point of user where incident occurred (the patient's full identity should not
be reported).(The contact point need not necessarily be the person who actually witnessed
the incident. It is recommended that health-care facilities have a contact person for all
incidents reported).
- manufacturer's preliminary comments;
- manufacturer's proposed next action, and timescale;
- a statement of whether the manufacturer is aware of similar incidents having an
impact on the current report
- if yes, the names of any other Competent Authorities to which these incidents have
been reported, and the reference/date of the report (s);
- any other EEA State in which the device is known to be on sale.
6.2.2 If the manufacturer is located outside the EEA, a suitable contact point within
the EEA should be provided. This may be the manufacturer's authorised representative,
persons responsible for placing devices on the market or any other agent authorised to act
on their behalf for purposes relating to Medical Devices Vigilance.
6.2.3 The report should not be unduly delayed because of incomplete information.
If the initial report is made by means other than by letter post (eg telephone, fax),
it should be followed as soon as possible by a written confirmation.
6.2.4 The report may also include a statement to the effect that the report is made by
the manufacturer without prejudice and does not imply any admission of liability for the
incident or its consequences.
6.3 COMPETENT AUTHORITY ACTIONS ON RECEIPT
OF AN INITIAL REPORT FROM A MANUFACTURER
6.3.1 The Competent Authority should acknowledge
the receipt of the report to the sender.
6.3.2 The Competent Authority should record the report - this should involve
categorising the incident, for example:
- by date (of incident, receipt by manufacturer, receipt by Competent Authority);
- by outcome (death, injury or near incident);
- by manufacturer and model;
- by device kind, using appropriate nomenclature;
- by "coordinating" Competent Authority for this type of incident (if any -
see paragraph 7.2);
- by the date when the manufacturer's next action is due.
6.3.3 The Competent Authority should evaluate the report and intervene as appropriate,
in consultation with the manufacturer if practicable (see para 7).
6.4 COMPETENT AUTHORITY ACTIONS ON REPORTS
FROM USER OR OTHER SYSTEMS
6.4.1 A report which appears to meet the criteria
of para 5, received by a Competent Authority from a User Reporting system or other source,
should be copied by the Competent Authority to the manufacturer without delay. In doing
so, patient confidentiality should be maintained.
6.4.2 Once the manufacturer has been so informed, the subsequent procedure is the same,
as far as practicable, as that described in Section 7 onwards of these Guidelines.
7. PROCEDURE FOLLOWING THE INITIAL
REPORT
7.1 PRINCIPLES
7.1.1 The manufacturer normally performs the
investigation, while the Competent Authority monitors progress. The Competent Authority
may intervene, or initiate independent investigation if appropriate. This should be in
consultation with the manufacturer where practicable. (See paragraph 7.4).
7.1.2 In the case of incidents of, groups of incidents, or recalls involving more than
one Competent Authority, there may emerge a single coordinating Competent Authority. Most
communications should then be between the "coordinating" Competent Authority and
the manufacturer. (See paragraph 7.2).
7.1.3 It is possible that the action concerning an incident may be completed without
further investigation following the initial report.
Note: The above principles are generalised and do not take account of interventions
by judicial or other agencies.
7.2 CO-ORDINATION BETWEEN COMPETENT
AUTHORITIES
7.2.1 Initial reports are not normally
disseminated between Competent Authorities. In the case of initial reports which confirm
that incidents meet the criteria set out in paragraph 8.4.1, information should be
disseminated between Competent Authorities and to the Commission at this stage (see
paragraph 8.4). In the unusual event that an inital report is to be disseminated, the
Competent Authority should inform the manufacturer prior to issue.
However, in order to achieve the purpose of the Vigilance system, any report made by a
manufacturer to an individual Competent Authority should be accessible in confidence to
the other Competent Authorities on request.
7.2.2 Competent Authorities should determine a single coordinating Competent Authority
under the following circumstances:
* incidents of similar types occurring in more than one country within the EEA, and
which lead to corrective action;
* recalls conducted in more than one country within the EEA, whether or not a
reportable incident has occurred.
The following hierarchy should determine the coordinating Competent Authority, unless
otherwise agreed between Competent Authorities:
* the Competent Authority which received the first initial report concerning this type
of incident.
* the Competent Authority in the State where the manufacturer or his authorised
representative is situated.
* the Competent Authority in the State where the Notified Body which made the
attestation leading to CE-marking, is situated;
7.2.3 The coordinating Competent Authority has responsibility for:
* monitoring the investigation and coordinating contacts with the manufacturer on
behalf of other Competent Authorities;
* accessing the expertise of the relevant Notified Body and coordinating with other
Competent Authorities within the EEA;
* discussing with the manufacturer the principles, need and circumstances of corrective
actions to be taken within the EEA;
* disseminating details of incidents which meet the criteria set out in paragraph 8.4.1
to other Competent Authorities and the Commission, including confirming the names of other
States affected by recalls etc.
* receiving and pooling data and experience from other Competent Authorities;
7.2.4 Confirmation of a co-ordinating Competent Authority where applicable should be
provided on the Competent Authority Report (see paragraph 8.4.2) under "reason for
report" (see Appendix 8.).
7.2.5 Such an arrangement would not affect the rights of an individual Competent
Authority to perform its own monitoring or investigation, or to instigate action within
its Member State in accordance with the provisions of the relevant Directives.
7.2.6 COMMITTEE ON MEDICAL DEVICES
If similar reports are being investigated in different Member States, it is recommended
that the Commission should be informed, so that it can facilitate or support measures such
as those described above. At the initiative of the Commission or the Member State, the
matter may be discussed in the framework of the Committee on Medical Devices.
7.3 MANUFACTURER ROLE FOLLOWING THE
INITIAL REPORT
7.3.1 The manufacturer normally performs the
investigation following the initial report, keeping the Competent Authority informed of
progress as appropriate.
7.3.2 If the manufacturer is not able to perform the investigation of an incident then
he should inform the Competent Authority without delay.
7.4 COMPETENT AUTHORITY ROLE FOLLOWING THE
INITIAL REPORT
7.4.1 The Competent Authority normally monitors
the investigation being carried out by the manufacturer. However, the Competent Authority
may intervene at any time. Such intervention should be in consultation with the
manufacturer where practicable.
7.4.2 Aspects of the manufacturer's investigation which may be monitored include, for
example:-course, or direction the investigation is taking;
- conduct, or how the investigation is being carried out;
- progress, or how quickly the investigation is being carried out;
- outcome, or whether the results are satisfactory.
7.4.3 Facts which may be needed include, for example:
- the number of devices involved;
- the length of time they have been on the market;
- details of design changes which have been made.
7.4.4 Liaison may be needed with:
- Notified Bodies (involved in the attestation leading to the CE marking);
- users;
- other Competent Authorities;
- other independent bodies, test houses etc.
7.4.5 The Competent Authority should consider liaison with other (non-medical device)
Competent Authorities, for example if a medicinal product is involved.
7.4.6 The Competent Authority should take coordinating action to ensure an
investigation is carried out if several manufacturers are involved.
7.4.7 If the manufacturer cannot for any reason perform the investigation, then the
Competent Authority should ensure an investigation is carried out. The manufacturer should
be kept informed.
7.4.8 Competent Authorities may also monitor experience with the use of devices of the
same kind (for instance, all defibrillators or all syringes), but made by different
manufacturers. They may then be able to take measures applicable to all devices of that
kind. This could include, for example, initiating user education or suggesting
re-classification
8 OUTCOME OF AN INVESTIGATION, AND
FOLLOW-UP
8.1 PRINCIPLES
8.1.1 Normally, the manufacturer should take the
action necessary following the investigation, including consultation with the Competent
Authority and performing any recalls - see paragraph 8.2.
8.1.2 The Competent Authority may take any further action it deems appropriate,
consulting with the manufacturer where possible - see paragraph 8.3.
8.1.3 There should be a final report which is a written statement of the outcome of the
investigation and of any action. This is made by the manufacturer to the Competent
Authority. If the Competent Authority performs the investigation then the manufacturer
should be informed of the result - see paragraphs 8.2, 8.3 and 8.6.
8.2 MANUFACTURER ACTIONS
8.2.1 The manufacturer should make a final report to the relevant Competent Authority -
see also paragraphs 8.3 and 8.6.
A suggested format for the manufacturer's final report is in Appendix 3.
8.2.2 Outcomes may include, for example:
- no action;
- additional surveillance or follow-up of devices in use;
- dissemination of information to users, eg by advisory notice;
- corrective action on future production;
- corrective action on devices in use;
- recall.
8.3 COMPETENT AUTHORITY ACTIONS
8.3.1 The Competent Authority should receive the final report from the manufacturer
concluding the investigation -see paragraph 8.6.
8.3.2 Competent Authority actions should be in consultation with the manufacturer
wherever practicable.
8.3.3 The Competent Authority should consider the content and method of dissemination
of any advisory note, in consultation with the manufacturer and medical practitioner if
appropriate - see paragraph 8.5.
8.3.4 Other Competent Authority actions may include, for example:
- no action;
- gathering more information, for example by commissioning independent reports;
- making recommendations to manufacturers, for example to improve information provided
with the device;
- keeping the Commission and other Competent Authorities informed, for example on
recalls and other actions to be taken; the information may be in the format of a Competent
Authority Report (see paragraph 8.4.2), or similar.
- consulting with the relevant Notified Body on matters relating to the conformity
assessment;
- consulting the Commission, for example if it is considered that re-classification of
the device is necessary;
- further user education;
- further recommendations to users;
- any other action to supplement manufacturer action.
8.3.5 The Competent Authority may take action in accordance with Article 2 of either
Directive, or in accordance with Article 7 of the AIMD or Article 8 of the MD.
8.3.6 The Competent Authority should consider whether action needs to be taken on
similar devices, made by the same or different manufacturer.
8.4 DISSEMINATION OF INFORMATION BETWEEN
COMPETENT AUTHORITIES
8.4.1 Information should be disseminated between
Competent Authorities and copied to the Commission for incidents where:
* corrective action (including recalls) is to be taken;
* there is a serious risk to the safety of patients or other users, but where no
corrective action has yet been established although measures are under consideration, or
where there is not yet a final report from the manufacturer.;
8.4.2 A recommended format for dissemination of information, using a "Competent
Authority Report" and notes for completion of the report are given at Appendix 8. The
manufacturer's report may be circulated with the Competent Authority Report.
The appropriate "reason for report" should be identified on the Competent
Authority Report. Competent Authorities receiving reports should pay particular attention
to the "reason for report" and any "recommendations" given by the
Competent Authority issuing the report. A number of reports may not require any immediate
further action. Wherever possible, Competent Authorities should direct enquiries arising
from the report to the Competent Authority providing the notification, who will
co-ordinate communication with the manufacturer or Notified Body.
8.4 3 Competent Authority Reports are intended for dissemination between Competent
Authorities and the Commission only, and are not for onward distribution to users or
other interested parties..
8.4.4 Competent Authorities should consult the manufacturer when preparing a report,
and should inform the manufacturer when one is issued.
8.5 DISSEMINATION OF INFORMATION OUTSIDE
COMPETENT AUTHORITIES
8.5.1 Careful consideration should be given to the
drafting and the dissemination of information by the Competent Authorities. The possible
positive and negative effects of the information to be disseminated should be considered
when drafting advisory notifications and when selecting the means and medium by which the
message is transmitted.
Preference should be given to notification directly to medical practitioner or
health-care facilities concerned.
Medical practitioners or other health-care professionals should be consulted where
appropriate.
The manufacturer should be consulted wherever practicable.
8.5.2 In exceptional circumstances, and only if other means are not appropriate,
dissemination of information direct to the public may be needed. The purpose of such
communication will normally be to suggest that patients or users contact their medical
practitioner for further, more specific advice
8.5.3 Consideration should be given to the preparation of a statement to the press for
use by all Competent Authorities.
8.5.4. The above considerations apply also to dissemination of information by the
manufacturer in consultation with the Competent Authorities.
8.5.5 Interfaces with communication media should be coordinated wherever practicable
between the manufacturer and the Competent Authorities.
8.6 SAFEGUARD CLAUSE
The application of the Vigilance system does not affect the responsibilities of the
Member States laid down in the Safeguard Clause (Article 7 of AIMD and Article 8 of MDD).
The Safeguard Clause procedures remain applicable regardless of the Medical Devices
Vigilance system.
8.7 COMPLETION OF THE INVESTIGATION AND
CLOSURE OF THE FILE
8.7.1 The Competent Authority should place the
manufacturer's final report on file and make any other observations necessary. The file
may then be endorsed as "closed".
8.7.2 If a Competent Authority itself conducts an investigation, the manufacturer (and,
where appropriate, other Competent Authorities) should be informed of progress and of the
results.
8.7.4 The final report should also be copied to any Competent Authorities who were
informed of the initial report.
8.7.5 It is possible for a file to be "closed" with no further action after
the initial report of the incident.
8.7.6 The Competent Authority should inform the manufacturer when a file is
"closed".
8.7.7 Files where action is complete, or for which no further action is intended,
should be retained as it is possible that changing circumstances may cause the matter to
be re-opened.